Features
Part Number
|
ER-100-IR
|
ER-100-1290/1650-ER=40
|
ER-100-VIS
|
Wavelength Range
|
850 nm to 1650 nm
|
1290 nm to 1650 nm
|
400 nm to 700 nm
600 nm to 1000 nm2
|
Extinction Ratio Range
|
30dB for 850 nm to 1290 nm
35dB for 1290 nm to 1650 nm
|
40dB for 1290 nm to 1650 nm
|
30dB
|
Dynamic Range
|
47dB
|
47dB
|
40dB
|
Extinction Ratio Accuracy
|
±1dB
|
||
Extinction Ratio Resolution
|
0.01dB
|
||
Angular Accuracy3
|
± 0.5 degrees
|
||
Angular Resolution
|
0.3 degrees
|
||
Update Rate (Extinction Ratio)
|
2.7 Hz
|
||
Update Rate (Relative Power)
|
650 Hz
|
||
Input Optical Power4
|
50 μW to 1.0 mW
|
||
Communication Interface
|
RS232
|
||
Input Supply Voltage5
|
Universal 50/60 Hz 110/220V AC/DC adapter
|
||
Dimensions (with protective boot)
|
65 x 90 x 184 mm (2.5 x 3.5 x 7.25 inches)
|
||
Weight (with protective boot)
|
0.48 kg (1.06 lb)
|
||
Operating Temperature
|
-10° to 50°C (14 to 122°F)
|
||
Storage Temperature
|
-30° to +60°C (-22 to 140°F)
|
||
Storage Humidity
|
<90% RH non condensing
|
||
Regulatory Compliance
|
Complies with EMC (EN61326) and Safety Requirements (EN61010-1)
|
||
1 Tested at 23°C ± 2°C with a 1550 nm linearly polarized source plus a master 1550 nm polarization maintaining patch cord, after a 30 minute warm-up period.
2 Depending on whether a receptacle for 400 to 700 nm or for 600 to 1000 nm wavelengths is selected.
3 For FC style connectors with high tolerance keyway.
4 Without attenuator. Higher powers can be measured with an attenuator. When ordering an attenuator with >20dB attenuation, the wavelength should be specified. Note: Exposing the detector to power higher than +17dBm (50 mW) without attenuator for a short period of time (3 minutes) can damage the detector.
5 North American power cord provided as standard equipment. Power cords for Europe and the United Kingdom can be ordered separately.
|